cui .com date 01/18/2012 page 1 of 4 series: vskm-s5 series description: medical switching power supply model output voltage output current output power ripple and noise 1 ef ciency (vdc) max (a) max (w) typ (mvp-p) typ (%) vskm-s5-3r3u 3.3 1.25 4.2 30 66 vskm-s5-5u 5 1 5 30 72 vskm-s5-9u 9 0.55 5 30 74 VSKM-S5-12U 12 0.42 5 30 76 vskm-s5-15u 15 0.33 5 30 76 vskm-s5-24u 24 0.23 5.5 30 78 notes: 1. ripple and noise measured at 20 mhz bandwidth part number key vskm-s5 - xx u base number output voltage features ? up to 5.5 w continuous power ? compact board mount design ? universal input (85~264 vac / 110~370 vdc) ? single output from 3.3 to 24 v ? over voltage, over temperature, and short circuit protections ? full medical approvals ? efficiency up to 78% rohs
cui .com date 01/18/2012 page 2 of 4 cui inc series: vskm-s5 series description: medical switching power supply input parameter conditions/description min typ max units voltage 85 110 264 370 vac vdc frequency 47 63 hz input current at 110 vac at 230 vac 110 70 ma ma inrush current at 110 vac at 230 vac 10 20 a a external input fuse (recommended) slow blow, 250 v 1 a output parameter conditions/description min typ max units line regulation 0.5 % load regulation 10 ~ 100% 1 % temperature coef cient 0.02 %/c hold-up time at 230 vac 50 ms voltage set point accuracy 3.3 v output all other models 3 2 % % switching frequency 100 khz protections parameter conditions/description min typ max units over voltage protection diode clamp and chip lock up short circuit protection auto recovery with no damage from a short on any output over-temperature protection 150 c safety & compliance parameter conditions/description min typ max units isolation voltage primary to secondary (for 1 minute) 4,000 vac safety approvals iec 60601-1, en 60601-1, ul 60601-1 safety class class ii emi/emc en 55011 (level a), iec/en 61000-4-2 (level 4, 8kv/15kv), iec/en 61000-4-3, iec/en 61000-4-4 (level 4, 4kv), iec/en 61000-4-5 (level 4, 2kv/4kv) rohs compliant yes mtbf 25c 300,000 hrs environmental parameter conditions/description min typ max units operating temperature -25 70 c storage temperature -40 105 c case temperature 95 c operating humidity non-condensing 95 %
cui .com date 01/18/2012 page 3 of 4 cui inc series: vskm-s5 series description: medical switching power supply derating curves input voltage (vac) load (%) 100 80 60 50 40 20 70 85 110 125 210 225 240 264 ambient temperature (c) load (%) 100 80 60 50 40 20 -40 -30 -25 -20 -10 0 10 50 60 70 1. output power vs. ambient temperature 2. output power vs. input voltage +vo pin connections pin function 1 2 ac(n) 3 ac(l) 4-vo 1 2 3 4 2 1 3 4 1.800 0.400 0.500 20.32 0.800 45.72 12.70 10.16 recommended pcb footprint (top view) grid: 2.54 mm (0.10 inch) 25.40 1.000 45.72 [1.800] 10 . 1 6[ 0 .4 0 0 ] ? 1.20 [0.047] 20. 32 [ 0 . 80 0] 25 .40 [ 1.00 0 ] +0 . 50 -0 .0 0 +0 .0 20 -0 .0 00 50.80 [2.000 ] +0. 50 -0.00 +0. 020 -0.000 6.00 15.16 0.597 (typ) 0.236 2.000 50.80 1.00 0.039 (typ) mechanical drawing mechanical parameter conditions/description min typ max units dimensions 2.0 x 1.0 x 0.6 inch (50.8 x 25.4 x 15.2 mm) case material ul94v-0 weight 35 g units: mm [inches] tolerance: 0.5 [0.02] pin section tolerance: 0.10 [0.004]
date 01/18/2012 page 4 of 4 cui inc series: vskm-s5 series description: medical switching power supply cui offers a two (2) year limited warranty. complete warranty information is listed on our website. cui reserves the right to make changes to the product at any time without notice. information provided by cui is believed to b e accurate and reliable. however, no responsibility is assumed by cui for its use, nor for any infringements of patents or other rights of third parties which may result from its use . cui products are not authorized or warranted for use as critical components in equipment that requires an extremely high level of reliability. a critical component is any component of a life support device or system whose failure to perform can be reasonably expected to cause the failure of the life support device or system, or to affect its safety or effectiveness. headquarters 20050 sw 112th ave. tualatin, or 97062 800.275.4899 fax 503.612.2383 cui .com techsupport@cui.com / +v0 -v0 l 1 n & |